An implantable cardioverter defibrillator (AICD) is a device
that is implanted in the chest to monitor for and, if necessary,
correct episodes of rapid heartbeat. If the heartbeat gets
too fast (ventricular tachycardia), the AICD will stimulate
the heart to restore a normal rhythm (anti-tachycardia pacing).
In cases where the heartbeat is so rapid that the person may
die (ventricular fibrillation), the AICD will also give an
electric shock (defibrillation) to reset the heartbeat.
An AICD is similar in many respects to an artifical pacemaker,
which is another type of device that corrects an abnormal
heart rhythm. However, pacemakers are usually chosen to correct
a heart rhythm that is too slow (bradycardia), whereas AICDs
are used to correct a heart rhythm that is too fast (tachycardia).
And there are patients who need both bradycardia pacing and
anti-tachycardia pacing. In these patients, an AICD will be
used to pace the heart.
The AICD is implanted into the chest of the patient during
a minor surgical procedure (not open-heart surgery). A short
stay in the hospital is usually required and some patients
may need to take medications that help the heart maintain
a normal rhythm (antiarrhythmics). Once the AICD is in place,
it runs on batteries for about four to seven years, depending
on how often an electric shock is discharged. AICD batteries
will not run out unexpectedly. Physicians can detect when
the battery is running low during a routine office visit.
People with AICDs need to be careful in certain situations.
More information on the lifestyle considerations of living
with an AICD can be found by clicking on the following link:
Living with an Implantable Defibrillator.
What is an implantable cardioverter defibrillator
(AICD)?
An implantable cardioverter defibrillator (AICD) is a device
that is implanted in the chest to monitor for and, if necessary,
correct episodes of rapid heartbeat. The implantable cardioverter
defibrillator gets its name from the two functions that it
performs. First, the AICD sends small electrical charges to
the heart to reset it when it goes too fast. This
process of converting one rhythm or electrical pattern to
another is called cardioversion. Second, the AICD will send
stronger charges to reset the heart if it begins
quivering instead of beating. The act of stopping this potentially
fatal quivering of the heart (ventricular fibrillation) is
called defibrillation.
Although the main functions of the AICD are cardioversion and
defibrillation, it can also be programmed to do the following:
- Anti-tachycardia pacing. When an AICD senses a fast but
rhythmic heartbeat (tachycardia), it releases a series of
precisely timed low-intensity electrical pulses that gently
interrupt the heart and allow it to return to a slower pace.
Whereas both cardioversion and defibrillation involve shocks
that may feel like a sudden kick in the chest, these low-intensity
stimuli are generally not felt by the patient.
- Bradycardia pacing. Like an artificial pacemaker, the
AICD can sense an abnormally slow heartbeat (bradycardia)
and send small electrical signals to pace the heart until
it restores and maintains a normal heart rate.
Modern AICDs can be programmed for all of the above functions.
The AICD also records heart activity and can transmit this
information to the physician during a routine check, allowing
the physician to better diagnose and monitor the underlying
conditions causing the patients arrhythmia.

Who may benefit from an AICD?
A series of landmark studies confirmed that AICDs significantly
reduce the risk of sudden cardiac death due to arrhythmias.
In fact, some of the studies were ended earlier than scheduled
due to the superior benefit shown by the AICD in comparison
to drugs used alone.
Based on the results of these studies, an AICD may be recommended
for patients who have experienced any of the following:
- Previous heart attacks, with weakened functioning of the
left ventricle. The performance of the left ventricle is
expressed numerically as the left ventricular ejection fraction.
It represents the proportion of blood in the heart that
is pumped out with each beat. A normal range is between
55 and 75 percent. An ejection fraction below 40 percent
has been shown to increase the risk of sudden cardiac death.
In heart attack survivors with reduced ejection fractions,
it has been found that an AICD plus antiarrhythmic drugs
significantly lowers the risk of sudden cardiac death, as
compared to antiarrhythmics used alone.
- History of ventricular tachycardia (VT) or ventricular
fibrillation (VF). For these patients, AICDs have clearly
improved survival compared to antiarrhythmic drugs.
- Coronary artery disease. Patients with coronary artery
disease may have an underlying arrhythmia. Studies have
shown that, in patients with coronary artery disease who
received an AICD, cholesterol reducing drugs may have an
anti-arrhythmic effect that can reduce the recurrence of
ventricular tachycardia or ventricular fibrillation.
- Cardiac arrest.
Not
every patient needs an AICD. AICDs are generally not necessary
when an abnormal rhythm (arrhythmia):
- Is an isolated occurrence with no underlying disease
- Is caused by an electrolyte imbalance or drug overdose
- Occurred within the first 48 hours after a heart attack
- For patients having an abnormally slow heartbeat (bradycardia),
the preferred treatment is an artificial pacemaker.

How an AICD works
Similar in structure to a pacemaker, the AICD consists
of three parts:
- Generator
- Leads
- Electrodes
The generator is a small box, usually about 2 inches wide
and approximately 3 ounces in weight. Some generators are
even smaller, measuring 1 inch in diameter and weighing about
half an ounce. They are battery-powered, and most use lithium
batteries that need to be replaced every four to seven years.
The generator is responsible for generating the electric shock.
Attached to the generator are two leads, or wires, generally
made of platinum with an insulating coating of either silicone
or polyurethane. The leads carry the electric shock from the
generator.
At the tip of each lead is a tiny device called an electrode
that delivers the necessary electrical shock to the heart.
Thus, the electric shock is created by the generator, carried
by the leads and delivered by the electrodes to the heart.
The decision of where to put the leads depends on the needs
of the patient, as described below.
The heart has four chambers. The two lower chambers, or ventricles,
perform most of the pumping action. The right ventricle pumps
oxygen-poor blood that has returned from the body into the
lungs to get more oxygen. The left ventricle takes that oxygen-rich
blood and pumps it back out to the body. While they are pumping
the blood to different areas, the ventricles need to be synchronized
(pumping at the same time) for blood to circulate properly.
The two upper chambers of the heart are called the atria (or
atrium if referring to only one). They also pump blood, but
with less force than the ventricles. Blood returning from
the body to the heart enters the right atrium, which pumps
that blood into the right ventricle and out to the lungs where
it picks up a new supply of oxygen. Similarly, the left atrium
pumps oxygen-rich blood returning from the lungs into the
left ventricle, which in turn pumps the oxygen-rich blood
through the aorta and out to the rest of the body. In order
for the heart to function properly, the atria must not only
be synchronized with each other, but with the ventricles as
well.
Depending on the needs of the patient, the two leads of an
AICD can be placed on two of these chambers in a variety of
combinations depending on the needs of the patient. Currently,
there is great research interest in placing leads in both
the right and left ventricles for patients with heart failure.
By simultaneously pacing both ventricles, the AICD may restore
normal heart synchronization. Although results of this technique
appear promising, it is still considered to be experimental.

How invasive is an AICD?
Because surgery is needed to implant the AICD, this is
considered an invasive treatment option. However, AICD insertion
is considered minor surgery and can be performed in either
an operating room or an electrophysiology laboratory. Patients
may be asked to stop taking certain medications (e.g., anticoagulants)
for several days prior to the surgery. They will also be asked
to sign a consent form and to dress in a hospital gown for
the procedure. The patients heart rate and blood pressure
will be monitored during the implantation.
This insertion site will be cleaned, shaved and numbed with
the injection of a local anesthetic. A small cut (incision)
is made in the chest wall just below the collarbone. Another
incision is made in the vein just under the collarbone. The
wires of the AICD are passed through the vein and attached
to the inner surface of the heart. The other ends of the wires
are connected to the main box of the AICD, which is inserted
into the tissue under the collarbone and above the breast.
Once the AICD is implanted, the physician will test it several
times by causing the heart to fibrillate, making sure the
AICD responds properly. Because the patient is anesthetized,
the patient will not feel this test. The incision is then
closed by sutures (stitches), staples or surgical glue. The
entire procedure takes about an hour.
Immediately following the procedure, a chest x-ray will be
taken to confirm the proper placement of the wires in the
heart. Patients will rest for several hours and their vital
signs will be closely monitored. The AICDs programming
may be adjusted using a magnetic wand that is passed over
the chest, during which the patient will not feel anything
different.
After the initial operation, the physician may induce ventricular
fibrillation or ventricular tachycardia prior to the patients
discharge. This allows the physician to program the AICD for
maximum efficiency.
Depending on the patients age and overall health, a
short stay in the hospital is usually required following AICD
insertion. The physician will provide specific instructions
regarding the patients appropriate activity level immediately
following the procedure. In general, patients may be instructed
not to bathe or shower for at least five days after the procedure.
They should also avoid contact sports, heavy lifting or vigorous
exercise for several weeks, in order to avoid dislodging the
wires.
A follow-up visit at the physicians office is usually
scheduled for about two weeks after the surgery. At this visit,
patients can expect the physician to do the following:
- Remove any remaining sutures (stitches) or surgical staples.
- Check the site for any signs of infection.
- Offer further instructions for how to live with an AICD.
If there are no complications, complete recovery from the
procedure will take about four weeks. During that time,
the wires will firmly take hold where they were placed.

Is there any risk of complications with
an AICD?
Although the insertion of an AICD requires only minor surgery,
it still carries some risks. While complications are rare,
patients should report any of the following symptoms immediately:
- Redness, warmth, tenderness or swelling of the incision
site, alone or with a fever. Sometimes a hard ridge forms
where the incision was closed. This will fade away as the
wound heals.
- Drainage of liquid from the incision site, alone or with
a fever.
- Increased shortness of breath, prolonged hiccuping or
difficulty breathing.
- Fainting, lightheadedness or dizziness.
- Fast or pounding heartbeats (palpitations).
- Chest pain.
- Re-experiencing the same symptoms that they had before
surgery.
Serious complications from the surgery occur in less than
1 percent of cases. These include:
- Severe bruising or bleeding
- Formation of a blood clot
- Torn blood vessel
- Punctured lung or heart muscle
- Stroke
- Heart attack
- Introduction of air into the space between the lung and
chest wall
- Death
The risk of having one of these complications is increased
if people have certain characteristics, such as the following:
- Advanced age
- Obesity (more than 20 pounds heavier than ones ideal
weight or body mass index 30 or greater)
- Severe lung disease (often due to smoking)
- Use of various medications
- Severely decreased heart function
Rarely, there may also be some complications with the AICD
itself. As with any mechanical device, the AICD or the wires
may malfunction. A small percentage of patients report at
least one time in which the AICD delivered a shock when no
arrhythmia was present. If patients feel a shock from the
AICD, they should call their physician. If other symptoms occur
such as dizziness, clamminess, palpitations, angina (chest
pain), loss of consciousness or blackouts, patients are advised
to call an ambulance and go to the emergency room, because
their AICD may need reprogramming.
Some patients may sustain a cluster of episodes of ventricular
tachycardia and/or ventricular fibrillation that results in
multiple shocks (discharges) being administered by the AICD.
This is known as an electrical storm. An electrical
storm is when at least three shocks and/or anti-tachycardia
pacing signals are delivered within 24 hours. Compared to
AICD patients not having such an episode, AICD patients surviving
an electrical storm have an increased risk of death, particularly
within the first three months following the event. What precisely
causes electrical storms is not known, but patients experiencing
them tend to have ventricular tachycardia and an ejection
fraction below 30 percent. They are also less likely to have
undergone revascularization (e.g., coronary bypass). Patients
who feel multiple discharges in a short time period should
contact their physician, who will review the data recorded
by the AICD and consider appropriate treatment.

How long will an AICD last?
Since the battery of the AICD is sealed within the main
box of the AICD, that entire box must be replaced when the
battery is low. Most modern AICDs use lithium batteries that
need to be replaced every four to seven years, depending on
how often an electric shock is discharged.
AICD batteries will not run out unexpectedly. When a battery
is running low, the elective replacement indicator (ERI) is
activated. Physicians can detect this activation during a
routine office visit. AICDs will continue to function for approximately
six months after the ERI is activated, allowing plenty of
time to schedule an elective replacement procedure.
Aside from the ERI, changes may occur in the way the AICD operates
that will indicate to a physician that the battery is beginning
to run low. Patients may or may not be able to feel these
changes in function. Regular communication with ones
physician is an important part of a successful experience
with an AICD.
